Coated VICRYL Polyglactin 910 Suture VIO 36IN(90CM) USP0(M3.5) S/A CT, Product Number J358H

FDA Recall #Z-0319-2025 — Class II — September 25, 2024

Recall #Z-0319-2025 Date: September 25, 2024 Classification: Class II Status: Ongoing

Product Description

Coated VICRYL Polyglactin 910 Suture VIO 36IN(90CM) USP0(M3.5) S/A CT, Product Number J358H

Reason for Recall

Complaints were received regarding needle pull-off and suture degradation. Analysis on returned product from this lot confirmed that some sutures were visually degraded in the package.

Recalling Firm

Ethicon, Inc. — Raritan, NJ

Classification

Class II — May cause temporary or medically reversible adverse health consequences.

Product Type

Devices

Product Quantity

31,032 units

Distribution

International distribution to the country of China.

Code Information

UDI-DI: 10705031237308 (each); 30705031237302 (box); Lot Number: QP2AJM

Status

Ongoing

Voluntary / Mandated

Voluntary: Firm initiated