VACUETTE Safety Blood Collection Set + Luer Adapter, 23x12" Safety Blood Collect LA Product, Item...

FDA Device Recall #Z-0835-2013 — Class II — February 1, 2013

Recall Summary

Recall Number Z-0835-2013
Classification Class II — Moderate risk
Date Initiated February 1, 2013
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Greiner Bio-One North America, Inc.
Location Monroe, NC
Product Type Devices
Quantity 192,000 pieces (192 cases)

Product Description

VACUETTE Safety Blood Collection Set + Luer Adapter, 23x12" Safety Blood Collect LA Product, Item number 450096. Usage: The Safety Blood Collection Set is used in routine venipuncture procedures.

Reason for Recall

The VACUETTE Safety Blood Collection Set + Luer Adapter may display leaking (small droplets) during use due to a crack on the luer connector piece.

Distribution Pattern

US Nationwide Distribution in the states of FL, IL, KY, NY, NC, PA, OH, TX, VA

Lot / Code Information

Lot No. 12H27

Other Recalls from Greiner Bio-One North America, Inc.

Recall # Classification Product Date
Z-0571-2026 Class II Brand Name: Vacuette Product Name: 8 ml CAT Se... Aug 19, 2025
Z-0408-2026 Class II Greiner BIO-ONE VACUETTE 2 ml 9NC Coagulation s... Aug 8, 2025
Z-2375-2024 Class II VACUETTE¿ TUBE 8 ml CAT Serum Separator Clot Ac... May 14, 2024
Z-0733-2024 Class II VACUETTE TUBE 4 ml FX Sodium Fluoride/Potassium... Dec 12, 2023
Z-1583-2023 Class II greiner bio-one VACUETTE TUBE 6 ml K3E K3EDTA, ... Mar 31, 2023

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.