Brand Name: 900-00218 Sterile Cover 10-Pack Shipper Box Assembly Product Name: 900-00218 Sterile...
FDA Device Recall #Z-2519-2025 — Class II — July 31, 2025
Recall Summary
| Recall Number | Z-2519-2025 |
| Classification | Class II — Moderate risk |
| Date Initiated | July 31, 2025 |
| Status | Ongoing |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Lumicell, Inc. |
| Location | Newton, MA |
| Product Type | Devices |
| Quantity | 200 units (20 packs containing 10 units per pack) |
Product Description
Brand Name: 900-00218 Sterile Cover 10-Pack Shipper Box Assembly Product Name: 900-00218 Sterile Cover 10-Pack Shipper Box Assembly Model/Catalog Number: 900-00218 Sterile Cover 10-Pack Shipper Box Assembly Software Version: Not Applicable Product Description: Sterile Cover Component: Lumicell Direct Visualization System (DVS) / PMA # P230014
Reason for Recall
The potential of the elastic bands included in the kit detaching or fracturing during surgical procedures. As a result, these failures may pose a risk of retained foreign bodies in the surgical cavity and potentially leading to surgical site infections.
Distribution Pattern
US Nationwide distribution in the states of CA, FL, MA, NC, TX.
Lot / Code Information
Model/Catalog Number: 900-00218 UDI-DIs: Sterile Cover 10-Pack Shipper Box Assembly: (01)0 0860008 53781 2, 840-00204 Assy, Dispenser Carton Sterile Cover: (01)0 0860003 25251 2, 810-00210 Assy, Outer Tray Sterile Cover: (01) 0 0860003 2520 5, Lot numbers: 43930, 43931, 43932, 43933, 42756, 44349, 44349-01, 44349-02, 44349-03, 44349-04;
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.