Infant AirLife single-limb, continuous-flow circuit (Model AH132) containing AirLife Humidificati...
FDA Recall #Z-0343-2018 — Class I — November 29, 2014
Product Description
Infant AirLife single-limb, continuous-flow circuit (Model AH132) containing AirLife Humidification Chamber, Model Number AH290. The AH290 Humidification Chamber and Heated Breathing Circuit Kit(s) operate as part of a system for delivering health and humidified respirator gasses to patients requiring invasive and non-invasive ventilation.
Reason for Recall
Manufacturing error in humidification chamber may cause an overfill/overflow condition of the chamber. The may case a water back-up into the patient circuit resulting in an excessive amount of water entering the airway/lungs of the ventilated patient.
Recalling Firm
Vyaire Medical — Vernon Hills, IL
Classification
Class I — Reasonable probability of serious adverse health consequences or death.
Product Type
Devices
Product Quantity
16,670 units in total
Distribution
US Nationwide Distribution
Code Information
CM17H12 CM17H16 CM17H17 CM17H19 CM17H20 CM17J27 CM17J28 CM17J29 CM17J30 CM17K09 CM17K20 CM17K21 CM17K22 CM17K23
Status
Terminated
Voluntary / Mandated
Voluntary: Firm initiated
What Should You Do?
Stop using this medical device immediately and consult your healthcare provider. Contact the recalling firm for instructions on return or replacement. If you experienced an adverse event, inform your doctor.