Visaris Vision/Auto, ddRAura OTC APS, Diagnostic X-Ray System, Version 2.2 The Vision systems ...
FDA Device Recall #Z-1880-2018 — Class II — April 5, 2018
Recall Summary
| Recall Number | Z-1880-2018 |
| Classification | Class II — Moderate risk |
| Date Initiated | April 5, 2018 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Visaris DOO |
| Location | Belgrade |
| Product Type | Devices |
| Quantity | 8 devices (US), 14 devices (international) |
Product Description
Visaris Vision/Auto, ddRAura OTC APS, Diagnostic X-Ray System, Version 2.2 The Vision systems are intended for use in all areas of diagnostic radiography imaging and specifically at secondary and tertiary levels of medical care. Powerful positioning capability makes this system particularly suited for institutions with immobile or hardly mobile patients, such as trauma wards and emergency departments. The system s versatility also makes it suitable for unified diagnostic imaging services for medium size hospitals where this system can cover all diagnostic radiography needs. In general, Vision systems are intended to position the patient, x-ray source and digital detector in such a way as to expose only the desired patient anatomy to X-rays and capture the resulting x-ray image. They are then intended to process the acquired image to a diagnostic viewing level and export it in DICOM format through an option of several distribution media including film, paper, CD-DVD and digital PACS archives. The system is intended for routine, daily, dedicated use in radiographic diagnosis. It can be used in any type of specialized radiology wards or for general radiographic diagnostics
Reason for Recall
A failure of the tube arm suspension is possible while the tube arm height is being adjusted. During tube height adjustment there is a small probability of failure of both the primary and redundant holding mechanisms, causing uncontrollable descent of the tube arm and a potential risk of injury to the patient.
Distribution Pattern
Worldwide Distribution - US Distribution to the states of : FL, LA, MD, MS, NJ, and NY., and to the countries of : Czech Republic, Germany, Serbia, Switzerland, Taiwan, and United Kingdom.
Lot / Code Information
System SN VC16B-02002, Column SN BA-215-1800-862-1-30008; System SN VC15L-02002, Column SN BA-215-1800-862-1-30015; System SN VC15L-01002, Column SN BA-215-1800-862-1-30011; System SN VC15J-01002, Column SN BA-215-1500-792-10001; System SN VC15I-02001, Column SN BA-215-1800-862-1-30006
Other Recalls from Visaris DOO
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-1882-2018 | Class II | Visaris Vision/Auto, ddRAura OTC APS, Diagnosti... | Apr 5, 2018 |
| Z-1881-2018 | Class II | Visaris Vision/Auto, ddRAura OTC APS, Diagnosti... | Apr 5, 2018 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.