Medtronic SynchroMed II, Model 8637, (The SynchroMed II Pump is supplied in 20 ml or 40 ml reserv...

FDA Device Recall #Z-1579-2013 — Class I — June 3, 2013

Recall Summary

Recall Number Z-1579-2013
Classification Class I — Serious risk
Date Initiated June 3, 2013
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Medtronic Neuromodulation
Location Minneapolis, MN
Product Type Devices
Quantity 261,109 total devices estimated implanted Worldwide

Product Description

Medtronic SynchroMed II, Model 8637, (The SynchroMed II Pump is supplied in 20 ml or 40 ml reservoir size.) Sterilized using ethylene oxide. The pump is part of an infusion system that stores and delivers a prescribed drug to a specific site. The implantable components of the SynchroMed II infusion system include the pump, catheter,r and catheter accessories The SynchroMed II Programmable Pump is indicated when patient therapy requires the chronic infusion of drugs or fluids:

Reason for Recall

Medtronic Neuromodulation is providing Healthcare Providers with safety information and patient management recommendations related to the potential for electrical shorting internal to they SynchroMed infusion pump. An electrical short circuit in a feedthrough may present as a motor stall or low battery reset/alarm and lead to a loss of or reduction in therapy which may result in the return of und

Distribution Pattern

worldwide

Lot / Code Information

all SynchroMed II pumps

Other Recalls from Medtronic Neuromodulation

Recall # Classification Product Date
Z-1043-2026 Class II A71200 Vanta" Clinician Programmer Application ... Dec 3, 2025
Z-0996-2026 Class II Restore Clinician Programmer Application (CP Ap... Nov 20, 2025
Z-1704-2025 Class II Enhanced Verify Evaluation Handset (CFN HH90130FA) Apr 3, 2025
Z-1543-2025 Class II Stimulation RC Clinician Programmer Application... Mar 4, 2025
Z-1506-2025 Class II A820 myPTM Software Application associated with... Jan 9, 2025

Frequently Asked Questions

Contact your physician immediately if you use an electrical device or implant affected by this recall. For external devices (monitors, pumps), stop using the device and contact the manufacturer for a replacement. For implanted devices (pacemakers, ICDs), your cardiologist will assess whether you need device replacement surgery — the decision depends on how dependent you are on the device and the actual failure rate observed in the field. Do not attempt to repair or modify a recalled electrical medical device yourself. Monitor for any new or unusual symptoms and report them to your doctor.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.