ICU Medical Plum Solo Precision IV Pump, List Numbers: 400010401 (US), 400011001 (Philippines)

FDA Recall #Z-1120-2026 — Class II — December 15, 2025

Recall #Z-1120-2026 Date: December 15, 2025 Classification: Class II Status: Ongoing

Product Description

ICU Medical Plum Solo Precision IV Pump, List Numbers: 400010401 (US), 400011001 (Philippines)

Reason for Recall

Plum Solo and Duo Infusion pumps include a feature, that when selected, automatically flushes the downstream line after a piggyback therapy. The programmed flush volume is delivered from the primary line container at the piggyback therapy rate after the piggyback therapy is complete. In certain cases, as described below, the clinician may receive an Upper Hard Limit (UHL) or Lower Hard Limit (LHL) Violation message when programming a piggyback flush, which prevents the flush from being programmed.

Recalling Firm

ICU Medical, Inc. — Lake Forest, IL

Classification

Class II — May cause temporary or medically reversible adverse health consequences.

Product Type

Devices

Distribution

Worldwide distribution - US Nationwide and the country of Philippines.

Code Information

400010401 (US), Software Versions: 1.0.2; 400011001 (Philippines), Software Versions: 1.0.0

Status

Ongoing

Voluntary / Mandated

Voluntary: Firm initiated