1) Cytomics FC 500 With CXP Software BECKMAN COULTER 2) Cytomics FC 500 MPL With MXP Software...

FDA Device Recall #Z-0471-2019 — Class I — January 12, 2018

Recall Summary

Recall Number Z-0471-2019
Classification Class I — Serious risk
Date Initiated January 12, 2018
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Beckman Coulter Inc.
Location Brea, CA
Product Type Devices
Quantity 3115

Product Description

1) Cytomics FC 500 With CXP Software BECKMAN COULTER 2) Cytomics FC 500 MPL With MXP Software BECKMAN COULTER flow cytometer for in vitro diagnostics

Reason for Recall

A manufacturing defect on the amplifier board of the Flow Cytometers my cause signal loss and/or signal drifting resulting in absence of data or a population shift in the data plots. This can impact patient results for any application.

Distribution Pattern

Worldwide distribution. US Nationwide including Puerto Rico. OUS: FC 500: Japan, South Korea, China, Italy, Spain, United Kingdom, Germany, Bangladesh, Czech Republic, France, Pakistan, India, South Africa, Austria, Saudi Arabia, Algeria, Turkey, Belgium, Netherlands, Bulgaria, Sweden, Australia, Hungary, Sri Lanka, Uruguay, Canada, New Zealand, Switzerland, Gabon, Portugal, Mexico, Singapore, Monaco, Hong Kong, Guadeloupe, Zambia, United Arab Emirates, Ireland, Reunion, Poland, Egypt, Slovakia, Croatia, Brazil, Chile, Bahrain, Slovenia, Bosnia and Herzegovina, Greece, Myanmar, Morocco, Serbia, Philippines, Lebanon, Romania, Moldova, Qatar, Barbados, Iraq, Oman, Namibia, Indonesia, Thailand, The Bahamas, Macau, El Salvador, Panama, Denmark, Venezuela, Azerbaijan, Israel, Colombia, Peru, Malaysia, Vietnam, Tunisia, Kuwait, Kenya, Libya, Latvia, Russia, Belarus, Ukraine, Kazakhstan, Taiwan EPICS XL: Albania, Algeria, Angola, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Botswana, Bulgaria, Burundi, Cameroon, Canada, China, Congo, Croatia, Cyprus, Czech Republic, Denmark, Dominican, Republic, Egypt, Ethiopia, France, Germany, Greece, Guinea, Guinea-Bissau, Hong Kong, India, Indonesia, Ireland, Israel, Italy, Japan, Kazakhstan, Kenya, Kuwait, Lebanon, Lesotho, Libya, Macau, Malawi, Mexico, Morocco, Mozambique, Namibia, Netherlands, New Zealand, Nigeria, Oman, Pakistan, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Singapore, Slovakia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Swaziland, Sweden, Switzerland, Syria, Taiwan, Tanzania, The Bahamas, Tunisia, Turkey, Turkmenistan, Uganda, Ukraine, United Arab Emirates, United Kingdom, Zambia, Zimbabwe

Lot / Code Information

FC 500 Catalog Num FC 500" Flow Cytometer MCL with UPS 626553 FC 500" Flow Cytometer MPL 626554 FC 500" MCL Flow Cytometer 5 Color (100V) 6605627 FC 500" MCL Flow Cytometer 5 Color (120V) 6605628 FC 500" MCL Flow Cytometer 5 Color (220V) 6605629 FC 500" MCL Flow Cytometer 5 Color (240V) 6605630 FC 500 Flow Cytometer, Remfg Assy, 5 Clr (220V) A89264 FC 500" Flow Cytometer 100-240V, 50-60HZ, Remfg Assy A89151 FC 500 Recond MCL UPS Flow Cytometer 100-240V, 50-60HZ A88262 FC 500 Flow Cytometer 100-240V A88263 All products manufactured affected for the time period: FC-500 From 01/22/2002 to 01/31/2018 EPICS XL/XL-MCL From 11/01/1992 to 03/15/2013

Other Recalls from Beckman Coulter Inc.

Recall # Classification Product Date
Z-0541-2026 Class II DxFLEX Flow Cytometer Catalog Number C78500 UD... Sep 22, 2025
Z-0441-2026 Class II SYNCHRON Systems Direct LDL Cholesterol (LDLD) ... Sep 17, 2025
Z-0071-2026 Class II SYNCHRON Systems Phosphorus (PHOSm) Reagent RE... Sep 3, 2025
Z-2573-2025 Class II UniCel DxH 690T COULTER Cellular Analysis Syste... Aug 13, 2025
Z-2574-2025 Class II UniCel DxH 800 COULTER Cellular Analysis System... Aug 13, 2025

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.