cobas infinity central lab, Material Number 07154003001

FDA Recall #Z-0952-2023 — Class II — December 12, 2022

Recall #Z-0952-2023 Date: December 12, 2022 Classification: Class II Status: Ongoing

Product Description

cobas infinity central lab, Material Number 07154003001

Reason for Recall

A complaint investigation revealed that an incorrect behavior relating to the rejection of orders capability of the Host Connectivity Agent (HCA) could occur where the order received from the Laboratory Information System (LIS) is created with an Internal ID rather than the External ID sent from the LIS. The erroneous event occurs under rare circumstances where the date in the sample's barcode does not match the order date sent from the LIS, and can lead to the order being mismatched to an another patient's sample ID rather than the subject patient's sample ID. The software bug impacts multiple cobas infinity central lab software (versions 2.5.x , 3.01.x, 3.02.x, and 3.03.x) and occurs when the afflicted software has been configured with specific pre-conditions.

Recalling Firm

Roche Diagnostics Operations, Inc. — Indianapolis, IN

Classification

Class II — May cause temporary or medically reversible adverse health consequences.

Product Type

Devices

Product Quantity

224 US; 2,797 worldwide

Distribution

Domestic distribution nationwide. Foreign distribution pending.

Code Information

cobas infinity central lab / cobas infinity core license using software versions (2.5.x , 3.01.x, 3.02.x, and 3.03.x) that have been configured with all of the following pre-conditions will be impacted: 1. Having Order ID configuration to include the date (Not Common in U.S. Configurations) 2. Having a host configured by Order rather than Tube (Not Common in U.S..Configurations) 3. Having message configured to receive registration date of the order using ExtRegisterDate and ExtRegisterHour. 4. Having HCA configuration enabled for " Reject orders if not possible to assign External ID" and the radio button option set to "To at least one test" UDI-DI 04015630936007

Status

Ongoing

Voluntary / Mandated

Voluntary: Firm initiated

What Should You Do?

Stop using this medical device immediately and consult your healthcare provider. Contact the recalling firm for instructions on return or replacement. If you experienced an adverse event, inform your doctor.

Back to All Device Recalls