LightMix¿ Zika rRT-PCR Test The LightMix¿ Zika rRT-PCR Test is a real-time RT-PCR test intended ...

FDA Device Recall #Z-0750-2017 — Class II — November 16, 2016

Recall Summary

Recall Number Z-0750-2017
Classification Class II — Moderate risk
Date Initiated November 16, 2016
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Roche Molecular Systems, Inc.
Location Branchburg, NJ
Product Type Devices
Quantity 24 pieces

Product Description

LightMix¿ Zika rRT-PCR Test The LightMix¿ Zika rRT-PCR Test is a real-time RT-PCR test intended for the qualitative detection of RNA from the Zika virus in serum or EDTA plasma from individuals meeting CDC Zika virus clinical criteria (e.g., clinical signs and symptoms associated with Zika virus infection) and/or CDC Zika virus epidemiological criteria (e.g., history of residence in or travel to a geographic region with active Zika transmission at the time of travel, or other epidemiologic criteria for which Zika virus testing may be indicated). Testing is limited to laboratories certified under the Clinical Laboratory Improvement Amendments of 1988 (CLIA), 42 U.S.C. ¿263a, to perform high complexity tests, or by similarly qualified non-U.S. laboratories. Results are for the identification of Zika virus RNA. Zika virus RNA is generally detectable in serum during the acute phase of infection (approximately 7 days following onset of symptoms, if present). Positive results are indicative of current infection. Laboratories are required to report all positive results to the appropriate public health authorities. Within the United States and its territories positive results must be reported to appropriate state/local public health authorities. Negative results do not preclude Zika virus infection and should not be used as the sole basis for patient management decisions. Negative results must be combined with clinical observations, patient history, and epidemiological information. The LightMix¿ Zika rRT-PCR Test is intended for use by trained clinical laboratory personnel specifically instructed and trained in the techniques of realtime PCR and in vitro diagnostic procedures. The Test is only for use under the Food and Drug Administration s Emergency Use Authorization.

Reason for Recall

The LightCycler¿ 480 algorithm used for the LightMix¿ Zika rRT-PCR Test, EUA (catalog number 07987897001) occasionally (~1% of the time) calls a positive result for a negative sample if there is an inflection point in the fluorescence signal.

Distribution Pattern

Nationwide Distribution

Lot / Code Information

38161611

Other Recalls from Roche Molecular Systems, Inc.

Recall # Classification Product Date
Z-1104-2023 Class II cobas SARS-CoV-2 & Influenza A/B Qualitative as... Dec 21, 2022
Z-1106-2023 Class II cobas Influenza A/B & RSV UC (Utility Channel) ... Dec 21, 2022
Z-1105-2023 Class II cobas SARS-CoV-2 & Influenza A/B Qualitative nu... Dec 21, 2022
Z-0097-2023 Class II cobas 5800 instrument, Material No. 08707464001... Sep 8, 2022
Z-0828-2022 Class II cobas LiatSystem, respiratory virus panel nucle... Nov 16, 2021

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.