Covidien Signia" Small Diameter Curved Tip Intelligent Reload, Model SIGSDS30CTVT

FDA Recall #Z-1458-2024 — Class II — February 14, 2024

Recall #Z-1458-2024 Date: February 14, 2024 Classification: Class II Status: Ongoing

Product Description

Covidien Signia" Small Diameter Curved Tip Intelligent Reload, Model SIGSDS30CTVT

Reason for Recall

Under certain firing conditions, reloads were found to articulate in an uncontrolled manner potentially causing disruption to the staple line. The issue is related to components in impacted reloads that are not fully secure, which leads to uncontrolled articulation. The issue can occur using either the Endo GIA" Ultra Universal Stapler or Signia" Stapling System.

Recalling Firm

Covidien, LP — North Haven, CT

Classification

Class II — May cause temporary or medically reversible adverse health consequences.

Product Type

Devices

Product Quantity

3,023 units

Distribution

Worldwide distribution - US Nationwide and the countries of France, Hong Kong, Israel and Japan.

Code Information

Model Number: SIGSDS30CTVT; UDI/DI: 10884521741888; Lot numbers: N3L1909UY, N3L1868UY, N3L2008UY, N3L2221UY, N3M1541UY, N3M1603UY, N3M1639UY;

Status

Ongoing

Voluntary / Mandated

Voluntary: Firm initiated