Phasitron 5, IPV In-Line Valve, (Intrapulmonary Percussive Ventilation) REF: P5-TEE-20 (pack of 2...
FDA Device Recall #Z-0031-2026 - Class I, September 17, 2025
Recall Summary
| Recall Number | Z-0031-2026 |
| Classification | Class I - Serious risk |
| Date Initiated | September 17, 2025 |
| Status | Ongoing |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Percussionaire Corporation |
| Location | Sandpoint, ID |
| Product Type | Devices |
| Quantity | 59160 |
Product Description
Phasitron 5, IPV In-Line Valve, (Intrapulmonary Percussive Ventilation) REF: P5-TEE-20 (pack of 20)/ P5-TEE (individual) with In-Line Valve Sentec Instructions for Use
Reason for Recall
IFU update; in-line valve, with pressure relief seal may leak when used with ventilator; don't use in neonates/infants, with target tidal volumes less-than50mL or less-than10kg- at high risk of cardiopulmonary/neurological compromise due to unrecognized hypoventilation, updates for continued use in others before design update; leak may cause respiratory acidosis, hypoxemia, or respiratory failure.
Distribution Pattern
Worldwide distribution. US nationwide, Russia, Japan, Switzerland, Canada, Qatar, Chile, South Africa, United Kingdom, Israel, Turkey, and UAE
Lot / Code Information
UDI-DI: 00849436000723. Lots: 240326, 250116, 240418, 250324, 240610, 250616, 240620, 241121, 230612, 240826, WO04884, 241118, WO06020, 241203, WO05019, 250111, WO04827, WO04756. Instructions for Use P20020 Rev F
Other Recalls from Percussionaire Corporation
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-0960-2025 | Class I | VDR4 Phasitron Breathing Circuit Models that co... | Dec 23, 2024 |
| Z-0208-2025 | Class I | Phasitron5 Breathing Circuit (P5-10, P5-HC, P5-... | Sep 16, 2024 |
| Z-3208-2024 | Class I | Percussionaire, Inc., REF: P5-TEE-20, Qty: 20,... | Aug 21, 2024 |
| Z-3207-2024 | Class I | Percussionaire, Inc., REF: P5-TEE, Qty: 1, Rx ... | Aug 21, 2024 |
| Z-0637-2024 | Class I | High Frequency Transport Phasitron Breathing Ci... | Dec 12, 2023 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic, removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency, Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.