Covidien Emprint Long Percutaneous Antenna with Thermosphere Technology, Material CA30L2 intended...

FDA Device Recall #Z-0625-2019 — Class II — November 15, 2018

Recall Summary

Recall Number Z-0625-2019
Classification Class II — Moderate risk
Date Initiated November 15, 2018
Status Ongoing
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Covidien LLC
Location North Haven, CT
Product Type Devices

Product Description

Covidien Emprint Long Percutaneous Antenna with Thermosphere Technology, Material CA30L2 intended for use in percutaneous, laparoscopic, and intraoperative coagulation (ablation) of soft tissue, including partial or complete ablation of nonresectable liver tumors.

Reason for Recall

The firm received consumer reports of the ceramic trocar tip of the Emprint ablation antenna disengaging from the needle shaft post-ablation. This can result in the tip remaining in the patient.

Distribution Pattern

US Nationwide

Lot / Code Information

GTIN Number - 10884521706606 Lot Numbers- S7HG007KX S7HG008KX S7HG009KX S7HG032LX S7HG033LX S7HG034LX S7JG008LX S7JG009X S7JG012X S7KG012LX S7KG038QX S7LG009X S7MG010LX S7MG019X S8BG004X S8BG005X S8CG002X S8DG001X S8DG026X S8EG004X

Other Recalls from Covidien LLC

Recall # Classification Product Date
Z-1557-2026 Class II HealthCast "Vital Sync" Remote Patient Monitori... Jan 30, 2026
Z-1775-2022 Class II Puritan Bennett 560 Ventilator, PB560, CFN 4096600 Aug 19, 2022
Z-1318-2022 Class I Palindrome SI Chronic Catheter, Sterile, Single... Jun 8, 2022
Z-1319-2022 Class I Palindrome HSI Chronic Catheter, Sterile, Singl... Jun 8, 2022
Z-1315-2022 Class I Palindrome Precision HSI Chronic Catheter, Ster... Jun 8, 2022

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.