T3F, Free T3 Assay, REF/Catalog Number L2KF32 (200 tests), Siemens Material Number (SMN) 10381675...

FDA Device Recall #Z-0868-2014 — Class II — August 27, 2013

Recall Summary

Recall Number Z-0868-2014
Classification Class II — Moderate risk
Date Initiated August 27, 2013
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Siemens Healthcare Diagnostics
Location Tarrytown, NY
Product Type Devices
Quantity 4695 kits in total

Product Description

T3F, Free T3 Assay, REF/Catalog Number L2KF32 (200 tests), Siemens Material Number (SMN) 10381675, and REF/Catalog Number L2KF36 (600 Tests), SMN 10381682; an IVD Immunoassay kit for use with the IMMULITE 2000/IMMULITE 200 XPi Analyzers --- Origin: UK Siemens Healthcare Diagnostics Products Ltd. Llanberis, Gwynedd, LL55 4EL UK. For the quantitative measurement of Free T3 in serum, as an aid in the clinical assessment of thyroid status.

Reason for Recall

Siemens Healthcare Diagnostics confirmed customer complaints regarding an increase in the number of euthyroid patients (those with normal function of the thyroid) demonstrating values above the recommended normal range as published in the Instructions For Use for the IMMULITE/IMMULITE 1000 and/or IMMULITE 2000/IMMULITE 2000 XPi for several lots of Free T3 kits. A positive bias in quality control results was also observed, but the values may remain within the established ranges.

Distribution Pattern

Worldwide Distribution-USA (nationwide) and the countries of Albania, Algeria, Argentina, Australia, Austria, Bangladesh, Belgium, Bulgaria, Bosnia-Herzegovina, Brazil, Belarus, Canary Islands, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, France, Germany, Great Britain, Greece, Guatemala, Honduras, Hungary, Iceland, India, Indonesia, Ireland, Israel, Iran, Italy, Jordan, Republic of South Korea, Kuwait, Kazakhstan, Latvia, Lebanon, Libya, Lithuania, Macedonia, Mexico, Malaysia, Malta, Nepal, Netherlands, Norway, Panama, Peru, Philippines, Pakistan, Paraguay, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Spain, Sri Lanka, South Africa, Switzerland, Turkey, Thailand, Taiwan, Uruguay, Usbekistan, Venezuela, Vietnam, and United Arab Emirates.

Lot / Code Information

Lot Numbers 737, 738, 739, 740, 741, 742, 743

Other Recalls from Siemens Healthcare Diagnostics

Recall # Classification Product Date
Z-0827-2018 Class II BC Thrombin Reagent Kit Lot Number 46751; UDI -... Oct 17, 2017
Z-0122-2015 Class II IMMULITE /IMMULITE 1000 Systems COR Cortisol, R... Sep 19, 2014
Z-2715-2014 Class II ADVIA Centaur Immunoassay System (including ref... Aug 19, 2014
Z-2716-2014 Class II ADVIA Centaur XP Immunoassay System (including ... Aug 19, 2014
Z-2662-2014 Class II ADVIA Centaur Immunoassay System, Catalog Numbe... Aug 13, 2014

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.