Brand Name: Leksell GammaPlan Product Name: Radionuclide RT Treatment planning system Software ...

FDA Recall #Z-1340-2026 — Class II — December 12, 2025

Recall #Z-1340-2026 Date: December 12, 2025 Classification: Class II Status: Ongoing

Product Description

Brand Name: Leksell GammaPlan Product Name: Radionuclide RT Treatment planning system Software Version: 11,3, 11.4 Product Description: Leksell GammaPlan¿ is a computer-based system designed for Leksell Gamma Knife¿ treatment planning.

Reason for Recall

Due to an error when defining a new stereotactic reference that users fail to first change an already defined obsolete reference to a pre-plan reference. This may result in a treatment plan that potentially irradiates the incorrect patient location if the error is not detected by users.

Recalling Firm

Elekta, Inc. — Atlanta, GA

Classification

Class II — May cause temporary or medically reversible adverse health consequences.

Product Type

Devices

Product Quantity

N/A

Distribution

Worldwide - U.S. Nationwide distribution including in the states of AK, AZ, CA, CO, ST, FL, GA, HI, IA, IL, LA, MA, MI, MN. MO, NC, NJ, NM, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, WA, WI, and WV. The countries of Algeria, Argentina, Australia, Belarus, Belgium, Brazil, Bulgaria, Canada, China, Colombia, Croatia, Czech Republic, Dominican Republic, Ecuador, Egypt, El Salvador, France, Georgia, Germany, Hong Kong, India, Iran, Iraq, Italy, Japan, Jordan, Republic of Korea, Kuwait, Libyan Arab Jamahiriya, Lithuania, Malaysia, Mexico, Morocco, Netherlands, Norway, Pakistan, Peru, Poland, Romania, Russian Federation, Saudi Arabia, South Africa, Spain, Sweden, Taiwan, Thailand, Turkey, United Kingdom, and Vietnam.

Code Information

Lot Code: 07340048311878 (LGP 11.3) 07340201502136 (LGP 11.4)

Status

Ongoing

Voluntary / Mandated

Voluntary: Firm initiated