Nova Biomedical Prime Plus Analyzer -Intended to measure certain gases in blood, serum, plasma or...
FDA Device Recall #Z-0207-2024 — Class II — September 18, 2023
Recall Summary
| Recall Number | Z-0207-2024 |
| Classification | Class II — Moderate risk |
| Date Initiated | September 18, 2023 |
| Status | Ongoing |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Nova Biomedical Corporation |
| Location | Waltham, MA |
| Product Type | Devices |
| Quantity | 1779 units: 438 US; 1341 OUS |
Product Description
Nova Biomedical Prime Plus Analyzer -Intended to measure certain gases in blood, serum, plasma or pH of blood, serum, and plasma. (1) PN 57400 Stat Profile Prime Plus Analyzer (2) PN 59508 Stat Profile Prime Plus Analyzer, Remanufactured
Reason for Recall
Potential risk of reporting erroneous creatinine patient results associated with the initial start-up of a new BUN/Creatinine Sensor Cartridge. The potential risk is within the first 2-4 hours after cartridge installation that may not be identified by traditional Quality Control practices.
Distribution Pattern
Nationwide Foreign: Country Algeria Australia Austria Bangladesh Benin Bolivia Brazil Canada Chile China Colombia Croatia CYPRUS Czech Republic Ecuador Egypt El Salvador Finland France Germany Greece Guatemala India Indonesia Ireland Israel Italy Japan JORDAN Kazakhstan Kenya Kurdistan, Iraq Kuwait Malaysia Malta Mexico Moldova Morocco Nepal Netherlands New Zealand Oman Pakistan Palestine Paraguay Paraquay Peru Philippines Poland Puerto Rico Qatar Republic of Panama Romania Russia Rwanda Saudi Arabia Serbia Slovenia South Africa South Korea Spain Sri Lanka Sudan Sweden Switzerland Syria Thailand Trinidad Trinidad,W.I. Tunisia UK Ukraine United Arab Emerites Uruguay Vietnam
Lot / Code Information
UDI-DI: (1) 00385480574006 (2) 00385480595087 All software versions (V1.17.1123.0 and V3.2.1123.0)
Other Recalls from Nova Biomedical Corporation
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-0682-2025 | Class I | StatStrip Glucose Ketone Hospital Meter System.... | Nov 20, 2024 |
| Z-0681-2025 | Class I | StatStrip Glucose Hospital Meter System. Model/... | Nov 20, 2024 |
| Z-0683-2025 | Class I | StatStrip Glucose Ketone (mmol/L) Hospital Mete... | Nov 20, 2024 |
| Z-0208-2024 | Class II | Nova Stat Profile Prime Plus POC Analyzer (CE)-... | Sep 18, 2023 |
| Z-2732-2017 | Class II | Nova Biomedical Stat Profile Prime Analyzer as ... | Jun 19, 2017 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.