Integra Licox Test Kit REF BC10R Rx Only An accessory provided for use with the LCXO2 and AC3...

FDA Device Recall #Z-1675-2014 — Class II — April 16, 2014

Recall Summary

Recall Number Z-1675-2014
Classification Class II — Moderate risk
Date Initiated April 16, 2014
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Integra LifeSciences Corp.
Location Plainsboro, NJ
Product Type Devices
Quantity 49 test probes

Product Description

Integra Licox Test Kit REF BC10R Rx Only An accessory provided for use with the LCXO2 and AC31 PtO2 Monitors. It provides users with a secondary means to verify the monitor's PtO2 state-of-calibration (but does not provide calibration) when needed by providing a specific electrical signal to the monitor.

Reason for Recall

Integra LifeSciences has identified through a small number of complaints that some lot numbers of Licox Test Kits (Catalog Number BC10R) may provide erroneous "calibration test failure" message when they are used with the Licox PtO2 Monitor (LCSO2, or may provide an erroneous out of range result with the Licox CMP Brain Tissue Oxygen and Temperature Monitor (AC31).

Distribution Pattern

Worldwide Distribution - USA including Illinois, Florida, Maryland, Massachusetts, North Carolina, Pennsylvania, South Dakota and Internationally to Australia, Austria, Belgium, Chile, Finland, Germany, Great Britain, Greece, Israel, Italy and Russia.

Lot / Code Information

Catalog Number BC10R Lot Nos: 051113, 100913, 140513, 161013, 25013

Other Recalls from Integra LifeSciences Corp.

Recall # Classification Product Date
Z-1295-2025 Class II AURORA Surgiscope System, Sterile, single use d... Feb 6, 2025
Z-1168-2025 Class II Brand Name: MediHoney¿ Gel Product Name: MediH... Jan 10, 2025
Z-0977-2025 Class II CODMAN CERTAS Plus Programmable Valve Inline Sm... Dec 16, 2024
Z-0978-2025 Class II CODMAN CERTAS Plus Programmable Valve Inline Sm... Dec 16, 2024
Z-0979-2025 Class II CODMAN CERTAS Plus Programmable Valve Right Ang... Dec 16, 2024

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.