Magellan Diagnostics LeadCare Ultra Blood Analyzer Part Number: 80-0010

FDA Recall #Z-1925-2017 — Class I — November 29, 2014

Recall #Z-1925-2017 Date: November 29, 2014 Classification: Class I Status: Terminated

Product Description

Magellan Diagnostics LeadCare Ultra Blood Analyzer Part Number: 80-0010

Reason for Recall

LeadCare Plus and LeadCare Ultra Blood Lead Testing Systems may generate falsely low lead concentrations in some blood samples. The issue is observed when blood samples are collected in tubes with lavender and tan rubber stoppers of the Becton Dickinson EDTA Vacutainer tubes, a collection device used with the testing system.

Recalling Firm

Magellan Diagnostics, Inc. — North Billerica, MA

Classification

Class I — Reasonable probability of serious adverse health consequences or death.

Product Type

Devices

Product Quantity

49 units

Distribution

Nationwide Distribution

Code Information

All serial numbers

Status

Terminated

Voluntary / Mandated

Voluntary: Firm initiated