GE Healthcare Discovery NM/CT 670 ES, Model/Catalog Numbers: 1) H3100XF ; System, Tomography, ...
FDA Recall #Z-0894-2025 — Class II — December 18, 2024
Product Description
GE Healthcare Discovery NM/CT 670 ES, Model/Catalog Numbers: 1) H3100XF ; System, Tomography, Computed, Emission
Reason for Recall
GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear Medicine systems (see Affected Products Details and Appendices) could have been transported or relocated without proper detector support, which can result in excessive load being applied to the detector mounting mechanisms. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
Recalling Firm
GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING — Tirat Carmel
Classification
Class II — May cause temporary or medically reversible adverse health consequences.
Product Type
Devices
Product Quantity
17 units
Distribution
Worldwide
Code Information
1) H3100XF: UDI/DI N/A, System ID NT1011, Serial Number T2AG12009/ 88227HL4; UDI/DI N/A, System ID NT2002, Serial Number T2AG12011/ 89154HL9; UDI/DI N/A, System ID NT1009, Serial Number T2AG12013/ 90088HL6; UDI/DI N/A, System ID NT1003, Serial Number T2BG11001/ 79573HL2; UDI/DI N/A, System ID NT1004, Serial Number T2AG12003/ 79574HL0; UDI/DI N/A, System ID NT1005, Serial Number 84775HL6; UDI/DI N/A, System ID NT1020, Serial Number T2AG12012/ 89666HL2; UDI/DI N/A, System ID NT1016, Serial Number T2AG13020/ 96237HL3; UDI/DI N/A, System ID NT0002, Serial Number 106491HL4; UDI/DI N/A, System ID NT2004, Serial Number 105724HL9; UDI/DI N/A, System ID NT1012, Serial Number T2AG13016/ 93885HL2; UDI/DI N/A, System ID NT1012, Serial Number T2AG13017/ 93885HL2; UDI/DI N/A, System ID NT1013, Serial Number 94165HL8; UDI/DI N/A, System ID NT1010, Serial Number T2AG12008/ 87697HL9; UDI/DI N/A, System ID NT3001, Serial Number 85748HL2; UDI/DI N/A, System ID NT1017, Serial Number T2AG13027/ 104213HL4; UDI/DI N/A, System ID NT1015, Serial Number T2AG13018/ 94292HL0.
Status
Ongoing
Voluntary / Mandated
Voluntary: Firm initiated