Pinnacle Hip Solutions ALTRX Polyethylene Acetabular Liner, Ref 1221-36-452 Product Usage: An...

FDA Device Recall #Z-1366-2013 — Class II — March 7, 2012

Recall Summary

Recall Number Z-1366-2013
Classification Class II — Moderate risk
Date Initiated March 7, 2012
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm DePuy Orthopaedics, Inc.
Location Warsaw, IN
Product Type Devices
Quantity 20 liners

Product Description

Pinnacle Hip Solutions ALTRX Polyethylene Acetabular Liner, Ref 1221-36-452 Product Usage: An acetabular liner is used in combination with other implants for total hip arthroplasty is intended to provide increased patient mobility and reduce pain by replacing the damaged hip joint articulation in patients where there is evidence of sufficient sound bone to seat and support the components.

Reason for Recall

DePuy Orthopaedics, Inc. (DePuy) is issuing a voluntary recall of one lot of the AltrX" liner because the product was incorrectly labeled on the box. The box was labeled for a 36IDX52OD +4 Neutral Liner, but the product enclosed was a 36IDX56OD Neutral Liner. Three ( 3 ) complaints were received in January 2012.

Distribution Pattern

Worldwide Distribution - USA Nationwide including the state of VA and the country of Ireland.

Lot / Code Information

product code1221-36-452, lot 187460.

Other Recalls from DePuy Orthopaedics, Inc.

Recall # Classification Product Date
Z-1456-2024 Class II ATTUNE AFFIXIUM Cementless Fixed Bearing Knee w... Mar 12, 2024
Z-1264-2023 Class II Attune Revision Limb Preservation System (LPS) ... Feb 15, 2023
Z-1267-2023 Class II Attune Posterior (PS) Fixed Bearing (FB) Tibial... Feb 15, 2023
Z-1263-2023 Class II Attune Revision Limb Preservation System (LPS) ... Feb 15, 2023
Z-1266-2023 Class II Attune Revision Limb Preservation System (LPS) ... Feb 15, 2023

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.