The Enzyme II Calibrator is an in vitro diagnostic product for the Alanine Aminotransferase (ALTI...

FDA Device Recall #Z-1217-2017 — Class III — May 5, 2016

Recall Summary

Recall Number Z-1217-2017
Classification Class III — Low risk
Date Initiated May 5, 2016
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Siemens Healthcare Diagnostics, Inc.
Location Newark, DE
Product Type Devices
Quantity 24,940.00

Product Description

The Enzyme II Calibrator is an in vitro diagnostic product for the Alanine Aminotransferase (ALTI) on the Dimension¿ clinical chemistry system. For prescription use only.

Reason for Recall

Complaints of positive shift of patient and QC results when the ALTI assay is calibrated with the Dimension Enzyme II calibrator. Internal investigation has confirmed that patient samples and QC material demonstrate a shift of up to 10% in comparison to results obtained using earlier calibrator lots.

Distribution Pattern

Distributed throughout the United States

Lot / Code Information

Product Lot: 4MD047, 5CD052, 5FD017, 5GD050, 5HD010, 5JD069, 5LD051, 6AD054

Other Recalls from Siemens Healthcare Diagnostics, Inc.

Recall # Classification Product Date
Z-1484-2026 Class II Atellica CH Urine Albumin (UAlb). Material Numb... Jan 20, 2026
Z-0491-2026 Class II IMMULITE 2000 Oak Mix Specific Allergen; Sieme... Oct 6, 2025
Z-0492-2026 Class II 3gAllergy Specific IgE Universal Kit; Kit Comp... Oct 6, 2025
Z-0493-2026 Class II Food Panel 3 Specific Allergen; Siemens Materi... Oct 6, 2025
Z-0366-2026 Class II Assay: IMMULITE 2000 Intact PTH; Test Code: iP... Sep 22, 2025

Frequently Asked Questions

Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.