HipLOC Hip Fracture Fixation System- Hiploc System 130¿ Plate 38 mm Barrel 4 Hole Plate Item Num...
FDA Device Recall #Z-3261-2018 — Class II — August 9, 2018
Recall Summary
| Recall Number | Z-3261-2018 |
| Classification | Class II — Moderate risk |
| Date Initiated | August 9, 2018 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Biomet UK Ltd. |
| Location | Bridgend |
| Product Type | Devices |
| Quantity | 1 unit |
Product Description
HipLOC Hip Fracture Fixation System- Hiploc System 130¿ Plate 38 mm Barrel 4 Hole Plate Item Number: 2110-130-004 Product Usage: The HipLOC Compression Hip Screw may be used primarily for fixation of bone fractures. Specific indications include open reduction and internal fixation of a wide variety of fractures of the proximal femur: intracapsular fractures, intertrochanteric fractures and subtrochanteric fractures
Reason for Recall
HipLOC System Plate Barrel due to an undersized bore on the HipLOC plate barrel, can result in the lag screw not fitting through the barrel
Distribution Pattern
US Nationwide in the states of AZ, MO, NJ and NM
Lot / Code Information
Lot Number: 3222008
Other Recalls from Biomet UK Ltd.
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-3262-2018 | Class II | HipLOC Hip Fracture Fixation System-Hiploc Syst... | Aug 9, 2018 |
| Z-3264-2018 | Class II | HipLOC Hip Fracture Fixation System-Hiploc Syst... | Aug 9, 2018 |
| Z-3263-2018 | Class II | HipLOC Hip Fracture Fixation System-Hiploc Syst... | Aug 9, 2018 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.