Medtronic Launcher Coronary Guide Catheter; MODEL: LA6JR40K
FDA Device Recall #Z-0205-2015 — Class II — October 13, 2014
Recall Summary
| Recall Number | Z-0205-2015 |
| Classification | Class II — Moderate risk |
| Date Initiated | October 13, 2014 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Medtronic Vascular |
| Location | Danvers, MA |
| Product Type | Devices |
| Quantity | 11 devices |
Product Description
Medtronic Launcher Coronary Guide Catheter; MODEL: LA6JR40K
Reason for Recall
Two reports have been received on the 6F launcher Guide Catheter lot (Model Number LA6JR40K; lot No 0007244836} for incorrect usable length of the Guide Catheter. It was reported that the catheters had a usable length of 90 cm instead of 55 cm, which was specified on the packaging label. Usable length is defined as the total length of catheter shaft from the distal end of the strain relief to the
Distribution Pattern
US Distribution including the states of NC, IA, IL and KS.
Lot / Code Information
Lot Number: 0007244836 MODEL: LA6JR40K
Other Recalls from Medtronic Vascular
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-1860-2021 | Class I | Angiographic Guidewire Component: Model Number:... | May 4, 2021 |
| Z-1245-2021 | Class II | Medtronic 6F Launcher Guide Catheter 100 cmJL 3... | Jan 15, 2021 |
| Z-1247-2021 | Class II | Medtronic 6F Launcher Guide Catheter 100cm EBU ... | Jan 15, 2021 |
| Z-1246-2021 | Class II | Medtronic 6F Launcher Guide Catheter 100 cm JR... | Jan 15, 2021 |
| Z-1248-2021 | Class II | Medtronic 6F Launcher Guide Catheter 100 cm EBU... | Jan 15, 2021 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.