Smith&Nephew ENGAGE Cementless Partial Knee System, Tibial Anchor Stem, Part Numbers: a) REF 1-...

FDA Recall #Z-1671-2023 — Class II — May 2, 2023

Recall #Z-1671-2023 Date: May 2, 2023 Classification: Class II Status: Ongoing

Product Description

Smith&Nephew ENGAGE Cementless Partial Knee System, Tibial Anchor Stem, Part Numbers: a) REF 1-10011-100 (SIZE 1-2), b) REF 1-10011-200 (SIZE 3-4), c) REF 1-10011-300 (SIZE 5-6); Unicondylar knee prothesis

Reason for Recall

Recent complaint data indicates that the revision rate may be trending higher than corresponding similar devices in global joint replacement registries. The data identifies a potential signal that the performance is an outlier versus the state of the art with respect to the risk for revision.

Recalling Firm

Smith & Nephew, Inc. — Memphis, TN

Classification

Class II — May cause temporary or medically reversible adverse health consequences.

Product Type

Devices

Product Quantity

1596 units

Distribution

US Nationwide distribution.

Code Information

a) REF 1-10011-100, UDI/DI 00885556873083; b) REF 1-10011-200, UDI/DI 00885556873090; c) REF 1-10011-300, UDI/DI 00885556873106; ALL BATCHES

Status

Ongoing

Voluntary / Mandated

Voluntary: Firm initiated