Knee Stem Extensions & Impactor Sleeve: STEM IMPLANT 20MMDX145MM STEM IMPLANT 22MMDX145MM STEM...

FDA Device Recall #Z-2552-2014 — Class II — July 22, 2014

Recall Summary

Recall Number Z-2552-2014
Classification Class II — Moderate risk
Date Initiated July 22, 2014
Status Terminated
Voluntary Voluntary: Firm initiated

Recalling Firm

Firm Zimmer, Inc.
Location Warsaw, IN
Product Type Devices
Quantity 972,125 units

Product Description

Knee Stem Extensions & Impactor Sleeve: STEM IMPLANT 20MMDX145MM STEM IMPLANT 22MMDX145MM STEM IMPLANT 24MMDX145MM SHARP FLUTED ST 20MMDX75MM SHARP FLUTED ST 22MMDX75MM SHARP FLTD ST 24MMDX75MM SHARP FLUTED ST 20MMDX130MM SHARP FLUTED ST 22MMDX130MM SHARP FLUTED ST 24MMDX130MM CEMENTED STEM EXT 13MM X 145MM IMPACTOR SLEEVE* NGK STEM EXT 12X245MM (200MM) NGK STEM EXT 13X245MM (200MM) NGK STEM EXT 14X245MM (200MM) NGK STEM EXT 15X245MM (200MM) NGK STEM EXT 16X245MM (200MM) NGK STEM EXT 17X245MM (200MM) NGK STEM EXT 18X245MM (200MM) NGK STEM EXT 20X245MM (200MM) NGK STEM EXT 22X245MM (200MM) SMOOTH REV STM 9.5MM X 19 SMOOTH REV STM 11.5MM X 1 SMOOTH REV STM 13.5MM X 1 SMOOTH REV STM 15.5MM X 1 SMOOTH REV STM 17.5MM X 1 REV FEM STM 10.5MM X 200M REV FEM STM 12.5MM X 200M REV FEM STM 14.5MM X 200M REV FEM STM 16.5MM X 200M 18.5X200MM REV STEM REV FEM STM 10.5MM X 250M REV FEM STM 12.5MM X 250M REV FEM STM 14.5MM X 250M REV FEM STM 16.5MM X 250M 12.5X155MM FLUTED OFFST R 14.5X155MM FLUTED OFFST R 16.5X155MM FLUTED OFFST R 18.5X155MM FLUTED OFFST R 20.5X155MM FLUTED OFFST R 12.5X175MM FLUTED REV STM FLUTE REV STEM - 14.5MM X FLUTE REV STEM - 16.5MM X FLUTE REV STEM - 18.5MM X FLUTE REV STEM - 20.5MM X FLUTE REV STEM - 22.5MM X FLUTED REV STM 10.5MM X 1 FLUTED REV STM 11.5MM X 1 FLUTED REV STM 12.5MM X 1 FLUTED REV STM 13.5MM X 1 FLUTED REV STM 14.5MM X 1 FLUTED REV STM 16.5MM X 1 FLUTED REV STM 18.5MM X 1

Reason for Recall

Zimmer is recalling sterile-packaged implants and instruments packaged in six different package configurations due to packaging design verification test failures. Specifically, multiple test samples from each of the six configurations failed simulated distribution and shipping testing when the devices either breached inner or outer package trays or seal integrity was lost.

Distribution Pattern

Worldwide Distribution.

Lot / Code Information

Item Numbers:00598801020, 00598801022, 00598801024, 00598801520, 00598801522, 00598801524, 00598801620, 00598801622, 00598801624, 00598801913, 31810604200, 32598801212, 32598801213, 32598801214, 32598801215, 32598801216, 32598801217, 32598801218, 32598801220, 32598801222, 621519095, 621519115, 621519135, 621519155, 621519175, 621520105, 621520125, 621520145, 621520165, 621520185, 621525105, 621525125, 621525145, 621525165, 681515125, 681515145, 681515165, 681515185, 681515205, 681517125, 681517145, 681517165, 681517185, 681517205, 681517225, 681519105, 681519115, 681519125, 681519135, 681519145, 681519165, 681519185 EXPIRATION DATE PRIOR TO 2024-07

Other Recalls from Zimmer, Inc.

Recall # Classification Product Date
Z-0981-2026 Class II Brand Name: Affixus Antegrade Femoral Nailing S... Dec 2, 2025
Z-0979-2026 Class II Brand Name: Affixus Antegrade Femoral Nailing S... Dec 2, 2025
Z-0973-2026 Class II Brand Name: Affixus Antegrade Femoral Nailing S... Dec 2, 2025
Z-0977-2026 Class II Brand Name: Affixus Antegrade Femoral Nailing S... Dec 2, 2025
Z-0967-2026 Class II Brand Name: Affixus Antegrade Femoral Nailing S... Dec 2, 2025

Frequently Asked Questions

Sterility recalls for medical devices vary in severity. If you have already had a procedure using a potentially non-sterile device, contact your healthcare provider immediately — you may need monitoring for signs of infection. Symptoms to watch for include fever, redness or swelling at the surgical site, unusual pain, or discharge. For devices that have not yet been used, they should be quarantined and returned to the manufacturer per the recall notice. Non-sterile implants can cause serious infections; early detection and treatment are critical.

Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.

Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.

What Should You Do?

Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.