Genius 2 Tympanic Thermometer Private Label Item Code: 3069
FDA Device Recall #Z-0994-2020 — Class II — October 28, 2019
Recall Summary
| Recall Number | Z-0994-2020 |
| Classification | Class II — Moderate risk |
| Date Initiated | October 28, 2019 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Cardinal Health 200, LLC |
| Location | Mansfield, MA |
| Product Type | Devices |
| Quantity | in total 37567 units US; 335573 OUS |
Product Description
Genius 2 Tympanic Thermometer Private Label Item Code: 3069
Reason for Recall
The frequency of calibration for the Genius Tympanic Thermometer as stated in the operating manual may not ensure that thermometers always remain within the stated accuracy range (¿ 0.2¿C for Genius 2 and ¿ 0.3¿C for Genius 3 thermometers). The measurement readings drift upwards over time, which means that the thermometers could exceed the upper stated accuracy tolerance of +0.2¿C for Genius 2 or +0.3¿C for Genius 3.
Distribution Pattern
Worldwide Distribution - US Nationwide and Foreign in the countries of: Australia Austria Belgium Canada Canary Islands Chile Costa Rica Croatia Cyprus Czech Republic Denmark Egypt Estonia Finland France French Guiana French Polynesia Gabon Germany Gibraltar Greece Guadeloupe Hong Kong Iran, Islamic Republic Of Ireland Israel Italy Kuwait Liechtenstein Luxembourg Martinique Mayotte Mexico Netherlands New Caledonia New Zealand Norway Panama Philippines Poland Portugal Puerto Rico Qatar Reunion Romania San Marino Saudi Arabia Singapore Slovakia Slovenia South Africa South Korea Spain Swaziland Sweden Switzerland Taiwan Thailand Turkey United Arab Emirates United Kingdom
Lot / Code Information
All product manufactured after October 1, 2016; Serial Numbers greater than N16601357
Other Recalls from Cardinal Health 200, LLC
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-1203-2026 | Class II | Chest Drainage Units and Accessories: Produc... | Dec 26, 2025 |
| Z-0897-2026 | Class II | ChemoPlus gowns: Product Code Product Descript... | Nov 12, 2025 |
| Z-0207-2026 | Class II | Salem Sump PVC Tubes: Product Code/Product N... | Sep 18, 2025 |
| Z-0208-2026 | Class II | Salem Sump PVC Tubes: Product Code/Product N... | Sep 18, 2025 |
| Z-0211-2026 | Class II | Salem Sump PVC Tubes: Product Code/Product N... | Sep 18, 2025 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic — removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency — Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.