PT113950 PT HYBRID Glenoid Post Regenerex porous titanium construct Sterile R Prosthesis, Sh...

FDA Recall #Z-0660-2013 — Class II — December 12, 2006

Recall #Z-0660-2013 Date: December 12, 2006 Classification: Class II Status: Terminated

Product Description

PT113950 PT HYBRID Glenoid Post Regenerex porous titanium construct Sterile R Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented

Reason for Recall

Biomet is recalling Part Number PT-113950 PT Hybrid Glenoid Post, following an investigation which identified that the male thread of the post may be oversized. This oversized condition can vary in degree and may lead to the following three events: 1) If the PT Hybrid Glenoid Post is not fully seated into the Hybrid Base and the implant construct is implanted, then a gap of 1-3 mm will be prese

Recalling Firm

Biomet, Inc. — Warsaw, IN

Classification

Class II — May cause temporary or medically reversible adverse health consequences.

Product Type

Devices

Product Quantity

1050

Distribution

Worldwide Distribution -- US (nationwide) including the countries of , Canada, Japan, Argentina, Australia ,Netherlands, Europe and Chile.

Code Information

Part 113950 Lot 093540, 093550,093570,093580,093590,093600,093610,093620,093720, 093730,093740,093760,093770,093780,282590,282600,282620,282630, 282680,282700,282710,489160,489180,489200,489210,489770,489780, 489790,714480,714550,714570,714700,714730,722210,722250,722280, 984770,984780,984790,984800,984810,984820,984830,984840, and 984850

Status

Terminated

Voluntary / Mandated

Voluntary: Firm initiated