MEDLINE UNITE FOOT & ANKLE Screw, Non-Locking, Mini T8 (2.0x8mm), REF MMSN2008
FDA Recall #Z-1045-2023 — Class II — December 22, 2022
Product Description
MEDLINE UNITE FOOT & ANKLE Screw, Non-Locking, Mini T8 (2.0x8mm), REF MMSN2008
Reason for Recall
The recall is due to observed intra-operative screw failure. When pre-drilled with ¿1.3mm drill bit for ¿2.0mm screws and ¿1.6mm drill bit for ¿2.4mm we have observed screw failure due to excessive torque.
Recalling Firm
MEDLINE INDUSTRIES, LP - Northfield — Northfield, IL
Classification
Class II — May cause temporary or medically reversible adverse health consequences.
Product Type
Devices
Product Quantity
420 units
Distribution
US Nationwide distribution.
Code Information
GTIN 10193489120387, Lot Number 810304
Status
Ongoing
Voluntary / Mandated
Voluntary: Firm initiated