Ulrich medical neon3 OCT spinal stabilization, REF CS 3917-12, Locking screw for crosslink, screw...

FDA Recall #Z-1350-2020 — Class II — December 31, 2019

Recall #Z-1350-2020 Date: December 31, 2019 Classification: Class II Status: Terminated

Product Description

Ulrich medical neon3 OCT spinal stabilization, REF CS 3917-12, Locking screw for crosslink, screw-to-screw

Reason for Recall

Due to a production error, the weld seam of the "Locking screw for crosslink, screw-to-screw" cannot reliably transmit the required torque of 3.5 Nm during installation.

Recalling Firm

ulrich medical USA Inc — Chesterfield, MO

Classification

Class II — May cause temporary or medically reversible adverse health consequences.

Product Type

Devices

Product Quantity

179 units

Distribution

AZ, TX, GA, FL, AR, CA

Code Information

Lot Numbers: U010577, U016620, U019084

Status

Terminated

Voluntary / Mandated

Voluntary: Firm initiated