EndoVive 3s Low Profile Balloon Kits Part Number: M00548750 (XMD P/N 70-0050-A18)
FDA Recall #Z-0977-2016 — Class II — December 23, 2015
Product Description
EndoVive 3s Low Profile Balloon Kits Part Number: M00548750 (XMD P/N 70-0050-A18)
Reason for Recall
A feeding (reflux) valve leakage issue. The feeding valve appears to close inconsistently, or close in a delayed manner, when an extension set is disconnected from the Low Profile Balloon device, thus allowing feeding solution or gastric contents to leak out of the device in some instances.
Recalling Firm
Xeridiem Mediem Medical Devices Inc — Tucson, AZ
Classification
Class II — May cause temporary or medically reversible adverse health consequences.
Product Type
Devices
Product Quantity
39
Distribution
Nationwide distribution to MA only.
Code Information
Part Number: M00548750 (XMD P/N 70-0050-A18) Lot Number: 929716
Status
Terminated
Voluntary / Mandated
Voluntary: Firm initiated