Moog Curlin Infusion, Ambulatory Infusion Pump The Curlin Ambulatory Infusion System includes C...

FDA Recall #Z-0688-2016 — Class II — December 23, 2015

Recall #Z-0688-2016 Date: December 23, 2015 Classification: Class II Status: Terminated

Product Description

Moog Curlin Infusion, Ambulatory Infusion Pump The Curlin Ambulatory Infusion System includes Curlin Ambulatory multi-therapy and PainSmart pumps. Curlin 4000, catalog number 340; Curlin 6000, 6000 CMS, and PainSmart IOD, catalog number 360; and Curlin Pump Calibration Set, catalog number 340-4038.

Reason for Recall

Moog is recalling Curlin 4000, 6000, and PainSmart Infusion Pumps manufactured or serviced by MMDG between March 18, 2015 and November 6, 2015 because the master pumps used to calibrate them were out of tolerance and may cause an over delivery of medications and solutions at a rate of 1.2 - 6.8% faster than programmed.

Recalling Firm

MOOG Medical Devices Group — Salt Lake City, UT

Classification

Class II — May cause temporary or medically reversible adverse health consequences.

Product Type

Devices

Product Quantity

11,472

Distribution

Worldwide Distribution - US (nationwide), Guam, and countries of: Australia, Canada, Croatia, Turkey, New Zealand, Saudi Arabia, Japan, Korea, and Italy.

Code Information

All manufactured or serviced by MMDG between March 18, 2015 and November 6, 2015. Curlin 4000, catalog number 340; Curlin 6000, 6000 CMS, and PainSmart IOD, catalog number 360; and Curlin Pump Calibration Set, catalog number 340-4038.

Status

Terminated

Voluntary / Mandated

Voluntary: Firm initiated