Cryptococcal Antigen Lateral Flow Assay (CrAg LFA), Reference # CR2003

FDA Recall #Z-1109-2017 — Class II — December 27, 2016

Recall #Z-1109-2017 Date: December 27, 2016 Classification: Class II Status: Terminated

Product Description

Cryptococcal Antigen Lateral Flow Assay (CrAg LFA), Reference # CR2003

Reason for Recall

The affected products have approximately a 91% specificity while the non-affected lots have approximately a 99% specificity. The recalling firm is aware of three customers who reported false positive results.

Recalling Firm

Immuno-Mycologics, Inc — Norman, OK

Classification

Class II — May cause temporary or medically reversible adverse health consequences.

Product Type

Devices

Product Quantity

630 units

Distribution

Nationwide throughout the US

Code Information

161028133653JH, 161003111544KH, and 161007105036JG

Status

Terminated

Voluntary / Mandated

Voluntary: Firm initiated