SpermMar Test 0.7ml Beads Particles Label on bottle: SpermMar Test IgA 0.7ml Beads/Particle...

FDA Recall #Z-1448-2015 — Class II — December 22, 2014

Recall #Z-1448-2015 Date: December 22, 2014 Classification: Class II Status: Terminated

Product Description

SpermMar Test 0.7ml Beads Particles Label on bottle: SpermMar Test IgA 0.7ml Beads/Particles-50 Tests IND REF SPMA_S LOT Expiration DO NOT FREEZE/NE PAS CONGELER FertiPro N.V. Label on box: SpermMar Test IGA 50 Determination Test Kit CONTENTS 1 vial SpermMar Latex Particles 0.7 ml REF SPMA_S FertiPro FertiPro N.V. Label on box: SpermMar Test IgA 50 Determination Test Kit Contents: 1 vial SpermMar latex particles, 0.7ml FertiPro N.V. For In Vitro Diagnostic Use SpermMar IqA Test Manufacturer's Product Number/Catalog Number: SPMA_S Lot/Serial Number Expiration Date FP14A09 02/29/2016 FP14A10 04/30/2016 FP14A11 04/30/2016 Product is a solution in filled into clear plastic bottles that is placed into a cardboard box. Both packages contain labeling information.

Reason for Recall

The firm is recalling three batches (FP14A09, FP14A10, FP14A11) of SpermMAR IgA test due to new formulation causing the motility of sperm cells to reduce in certain samples when the new bead solution was added.

Recalling Firm

Vitrolife Inc — San Diego, CA

Classification

Class II — May cause temporary or medically reversible adverse health consequences.

Product Type

Devices

Product Quantity

45 kits

Distribution

US Distribution to the state of : OR., and Internationally to Argentina.

Code Information

Manufacturer's Product Number/Catalog Number: SPMA_S Lot/Serial Number Expiration Date FP14A09 02/29/2016 FP14A10 04/30/2016 FP14A11 04/30/2016

Status

Terminated

Voluntary / Mandated

Voluntary: Firm initiated