30" (76 cm) Appx 3.7 ml, Ext Set w/Remv 2 Gang CLAVE¿ Stopcock, Remv CLAVE¿ Stopcock, NanoClave¿...
FDA Recall #Z-0698-2014 — Class II — December 24, 2013
Product Description
30" (76 cm) Appx 3.7 ml, Ext Set w/Remv 2 Gang CLAVE¿ Stopcock, Remv CLAVE¿ Stopcock, NanoClave¿ T-Connector, Rotating Luer, 1 Ext, Part No. A1000, Item No. SC124 The ICU Needleless Connector is a single use, sterile, non-pyrogenic device intended for use as an accessory to intravascular administration set for the administration of fluids to a patient through a cannula placed in the vein or artery.
Reason for Recall
ICU Medical is recalling certain lots of the NanoClave Needlefree Connector product line because of a small percentage of devices that may leak.
Recalling Firm
ICU Medical, Inc. — San Clemente, CA
Classification
Class II — May cause temporary or medically reversible adverse health consequences.
Product Type
Devices
Product Quantity
2250 units
Distribution
Worldwide Distribution - USA including UT, TX, GA, WI, VT, NH, AL, OH, TN, IL, CA, KY,HI, NC, and Internationally to CA, BE, IN, KW, and AU.
Code Information
Item No. SC124. Lot No. 2774006, 2778500, 2783068, 2794495.
Status
Terminated
Voluntary / Mandated
Voluntary: Firm initiated