BCS(R) Behring Coagulation System, Device catalog No. OVIO03 Product Usage: Multipurpose sys...

FDA Recall #Z-1867-2012 — Class II — December 1, 2010

Recall #Z-1867-2012 Date: December 1, 2010 Classification: Class II Status: Terminated

Product Description

BCS(R) Behring Coagulation System, Device catalog No. OVIO03 Product Usage: Multipurpose system for In Vitro coagulation studies

Reason for Recall

Firm has confirmed that if the Cleaner SCS vial becomes empty after processing a test for Antithrombin assay using the INNOVANCE(R) Antithrombin kit, and the test immediately following is an APTT based clotting assay several results may be shortened.

Recalling Firm

Siemens Healthcare Diagnostics, Inc. — Newark, DE

Classification

Class II — May cause temporary or medically reversible adverse health consequences.

Product Type

Devices

Product Quantity

186

Distribution

Worldwide Distribution - US (nationwide) including the states of: AZ, AR, CA, CO, CT, FL, GA, ID, IL, IN, IO, KS, KY, LA, MD, MA, MI, MS, MN, MT, NE, NH, NJ, NY, NC, ND, OH, OK, OR, PA, TN, TX, UT, VA, WA, WV, and WI and the countries of: Australia, France, Canary Islands, Spain, Belgium, Germany, Switzerland, Netherlands, Italy, Guadeloupe, Ceuta, Austria, Slovenia, Croatia, Saudi Arabia, Turkey, Greece, and San Marino.

Code Information

Device catalog No. OVIO03 (not lot specific)

Status

Terminated

Voluntary / Mandated

Voluntary: Firm initiated