HAMILTON-T1 Ventilator with software versions 1.1.2 or lower. The HAMILTON T-1 Ventilator i...

FDA Recall #Z-0741-2013 — Class I — December 19, 2012

Recall #Z-0741-2013 Date: December 19, 2012 Classification: Class I Status: Terminated

Product Description

HAMILTON-T1 Ventilator with software versions 1.1.2 or lower. The HAMILTON T-1 Ventilator is intended to provide positive pressure ventilatory support to adults and pediatrics

Reason for Recall

Analysis of a customer complaint has shown that during ventilation of small pediatric patients with high airway resistance and low lung compliance, the oxygen consumption of a Hamilton T1 ventilator with software versions 1.1.2 and lower must be calculated using a larger margin than originally expected.

Recalling Firm

Hamilton Medical Inc — Reno, NV

Classification

Class I — Reasonable probability of serious adverse health consequences or death.

Product Type

Devices

Product Quantity

97 total ventilators; 69 of part number 161005 and 28 of part number 161006

Distribution

US Nationwide Distribution including the country of Switzerland.

Code Information

Software Versions 1.1.2 and below; Part numbers 161005, 161006.

Status

Terminated

Voluntary / Mandated

Voluntary: Firm initiated