Software : myNeedle Guide 2D license, myNeedle Guide 2D AWP only Workflow license, or myNeedle...

FDA Recall #Z-1003-2023 — Class II — December 22, 2022

Recall #Z-1003-2023 Date: December 22, 2022 Classification: Class II Status: Ongoing

Product Description

Software : myNeedle Guide 2D license, myNeedle Guide 2D AWP only Workflow license, or myNeedle Guide 3D license. Used on the following CT (Computed tomography systems): SOMATOM go.Now Model 11061610 NA* SOMATOM go.Now Model 11061618 NA* SOMATOM go.Up Model 11061620 SOMATOM go.Up Model 11061628 NA* SOMATOM go.All Model 11061630 SOMATOM go.All Model 11061638 NA* SOMATOM go.Top Model 11061640 SOMATOM go.Top Model 11061648 NA* SOMATOM X.cite Model 11330001 SOMATOME X.ceed Model 11330002 SOMATOM go.Sim Model 11061660 NA* SOMATOM go.Open Pro Model 11061670 NAEOTOM Alpha Model 11330003

Reason for Recall

In the syngo.CT VA40 and syngo.CT VA50 myNeedle Guide application, the order of the orientation labels head , center, and feet in the axial thick slice segments are erroneously swapped

Recalling Firm

Siemens Medical Solutions USA, Inc — Malvern, PA

Classification

Class II — May cause temporary or medically reversible adverse health consequences.

Product Type

Devices

Product Quantity

29 units U.S.

Distribution

Nationwide

Code Information

UDI: 04056869151571, 04056869051314, 04056869151564, 04056869151571, 04056869231044 04056869231051, 04056869249247, 04056869263168, ¿04056869051314 S/N: Serial Number 112285 112443 117977 117982 119197 123059 123206 123222 123225 123233 123236 123261 123273 123281 123306 123310 123327 123329 125051 125067 127089 130121 172649 172758 172820 172914 172925 172949 172956

Status

Ongoing

Voluntary / Mandated

Voluntary: Firm initiated