Qualicheck 1+, Aqueous Quality Control, Level 2, S7140, REF 944-042. The solution is used for...

FDA Recall #Z-0608-2013 — Class III — June 6, 2006

Recall #Z-0608-2013 Date: June 6, 2006 Classification: Class III Status: Terminated

Product Description

Qualicheck 1+, Aqueous Quality Control, Level 2, S7140, REF 944-042. The solution is used for quality check of blood gas analyzers.

Reason for Recall

The barcode for the ABL77 on the insert incorrectly enters a control range for pH of 7.90-7.95 into the analyzer. The insert correctly states that the control range is 7.40-7.45.

Recalling Firm

Radiometer America Inc — Westlake, OH

Classification

Class III — Not likely to cause adverse health consequences.

Product Type

Devices

Product Quantity

840 Ampules

Distribution

Distributed to KY, NC, and NY.

Code Information

Lot R0040, expiration date February 2008; Lot R0042, expiration date April 2008.

Status

Terminated

Voluntary / Mandated

Voluntary: Firm initiated