Pointe Scientific CRP(HS) Wide Range Control Set and Pointe Scientific CRP(HS) Wide Range Standa...

FDA Recall #Z-2631-2017 — Class III — July 21, 2009

Recall #Z-2631-2017 Date: July 21, 2009 Classification: Class III Status: Terminated

Product Description

Pointe Scientific CRP(HS) Wide Range Control Set and Pointe Scientific CRP(HS) Wide Range Standard Set 5 Levels of 1 x 2 ml ready to use calibrator or standard The Pointe Scientific CRP (HS) Wide Range Multi-Calibrator Set is intended to be used for the calibration of the manufacturer's CRP (High Sensitivity) Wide Range immunoturbimidmetric assay. For in vitro diagnostic use only. The Pointe Scientific CRP (HS) Wide Range Multi-Standard Set Is intended to be used for the calibration of the manufacturer's CRP (High Sensitivity) Wide Range immunoturbimidmetric assay. For in vitro diagnostic use only

Reason for Recall

Individual vials of calibrator and standard were mislabeled; specifically the units on the vial labels are incorrectly listed as mg/dl when the correct unit is mg/L.

Recalling Firm

Medtest Holdings, Inc. — Canton, MI

Classification

Class III — Not likely to cause adverse health consequences.

Product Type

Devices

Product Quantity

323 vials

Distribution

Worldwide Distribution - US including CA, MS, NJ, PA, HI, MD, FL, OK, NV. Foreign Distribution to CHILE

Code Information

Model #'s C7568-CTL and C7568-STD Lots: C7568-STD: 717303-131. Exp 6/2009. C7568-CTL: 902701-111 and 902701-131. Exp 1/2011.

Status

Terminated

Voluntary / Mandated

Voluntary: Firm initiated