ISE Electrolyte Reference reagent (for Synchron LX Systems and UniCel DxC 600/800 Systems), Part ...
FDA Device Recall #Z-1951-2013 - Class II, September 22, 2010
Recall Summary
| Recall Number | Z-1951-2013 |
| Classification | Class II - Moderate risk |
| Date Initiated | September 22, 2010 |
| Status | Terminated |
| Voluntary | Voluntary: Firm initiated |
Recalling Firm
| Firm | Beckman Coulter Inc. |
| Location | Brea, CA |
| Product Type | Devices |
| Quantity | 718 units total (132 units in US) |
Product Description
ISE Electrolyte Reference reagent (for Synchron LX Systems and UniCel DxC 600/800 Systems), Part No. A28937, A28945, 467935, 467915. Product Usage: ISE Electrolyte Buffer reagent and ISE Electrolyte Reference reagent, when used in conjunction with SYNCHRON LX System(s), UniCel DxC 600/800 System(s) and SYNCHRON Systems AQUA CAL 1, 2 and 3, are intended for the quantitative determination of potassium concentration in human serum, plasma or urine.
Reason for Recall
Beckman Coulter is recalling the Synchron Systems LX20/UniCel DxC ISE Reference Reagent because it has a potassium (K) response near the instrument calibration flag limits and may exceed the limits on some instruments.
Distribution Pattern
Worldwide Distribution - USA Nationwide and the countries of Belgium, France, Germany, Netherlands, Poland, Spain, Sweden, Switzerland, and United Kingdom
Lot / Code Information
Lot No. T006042
Other Recalls from Beckman Coulter Inc.
| Recall # | Classification | Product | Date |
|---|---|---|---|
| Z-2973-2026 | Class II | Beckman Coulter DxC 500i Clinical Analyzer Modu... | Jul 15, 2026 |
| Z-2974-2026 | Class II | Beckman Coulter DxC 500 AU Clinical Analyzer Mo... | Jul 15, 2026 |
| Z-2888-2026 | Class II | REF # D23352, MicroScan LabPro V5.03 Panel Upda... | Jun 26, 2026 |
| Z-2889-2026 | Class II | REF # D23354, MicroScan LabPro V5.03 Panel Upda... | Jun 26, 2026 |
| Z-0541-2026 | Class II | DxFLEX Flow Cytometer Catalog Number C78500 UD... | Sep 22, 2025 |
Frequently Asked Questions
Contact your healthcare provider and the device manufacturer immediately. Check whether your specific model number and lot number are included in the recall scope. For external devices, stop using the affected product and arrange a replacement. For implanted devices, do not panic, removal is typically not required unless the risk assessment clearly indicates it. Your physician will guide you based on your individual clinical situation and the FDA's recommended actions. Report any adverse effects you may have experienced to FDA MedWatch at 1-800-FDA-1088.
Class I recalls indicate a reasonable probability of serious adverse health consequences or death from the defect. Class II recalls involve products that may cause temporary or medically reversible adverse health consequences, or where serious consequences are remote. Class III recalls cover products not likely to cause any adverse health consequences, typically involving technical regulatory violations. The classification guides urgency, Class I recalls require immediate action, while Class III may simply involve returning a product or acknowledging a labeling change. Always read the specific recall notice for recommended patient actions.
Report problems with medical devices to the FDA through MedWatch at 1-800-FDA-1088 or online at FDA.gov/safety/medwatch. Healthcare facilities are required by law to report device-related serious injuries and deaths. Patients and consumers can also report voluntarily. Include the device name, manufacturer, model number, and a description of the problem and any patient outcome. Reports from patients and clinicians help the FDA identify emerging safety signals and may trigger investigations that lead to recalls of dangerous devices.
What Should You Do?
Stop using this device if you are affected by this recall. Contact your healthcare provider and the manufacturer immediately for guidance. Report adverse events to FDA MedWatch.